- Hand-held instrument in ATEX EEx ia IIC T4 type of protection to which all Höntzsch standard vane wheel probes, including those sensing ± flow direction and with integrated temperature probe, can be connected.
- Input (8-pin flange socket) software commutable for connection of designated type of sensor.
- High-contrast LCD liquid crystal display, measured value display 13 mm high four-digit with decimal point as well as affiliated display unit, display of duration of measurement in seconds 6 mm high, adjustment and configuration instructions, operator's information, status displays 3 mm high.
- Interface RS232 for direct transfer of the stored measured values to a PC.
- Instantaneous values every second. Display of actual velocity v, actual volume flow V/t or temperature (resolution 0.1 °C), selectable.
- Display of instantaneous value in the START and also START/STOP mode also with long-term measurement. Display units: m/s, m³/h. l/min, °C .
- Power supply by Li-Ion accumulator. Charge by using the appropriate charger in the 6-pin flange socket. Charging time approx. 4 hours, running time approx. 50 hours.
- Analogue output
- FA: 0 . 4 V = 0 . x m/sFAR: 0 . 2 . 4
- V= - x . 0 . + x m/s, x = variable
- Data and measured value storage for 500 data records. When being stored the data records are automatically marked with a group number (max. 99) and serial number.
|